We believe this site might serve you best::

United States

United States

Choose language: English

Promega's Cookie Policy

Our website uses functional cookies that do not collect any personal information or track your browsing activity. When you select your country, you agree that we can place these functional cookies on your device.

GloMax®-Multi+ with Instinct Software Outdated Versions

  • Print
  • Email

Download GloMax Multi+ Detection System Firmware (version 04.11.0) and
Software
(version 2.2.0) (.zip)

To perform the update, follow these steps:

  1. Copy the firmware files (af.ldf, ff.ldf and fc.ldf) provided to the root of the USB stick and insert into the USB drive on the instrument.
  2. Press the Ok button when asked to update firmware.
  3. Remove the firmware files from the USB stick after process is complete.
  4. Copy the executable file (*.exe) provided to the root of the USB stick and insert into the USB drive on the instrument.
  5. Press the Ok button when asked to update software.
  6. Remove executable file from the USB stick after process is complete.
  7. Run empty plate for each detection module as well as for each fluorescence optical kit immediately after SW/FW upgrade. It is not necessary to read the entire plate. Reading one well will be sufficient (e.g, select well A1 only).

Note: The empty plate runs will take longer than current reads and once this step is completed for each optical kit and module the normal expected plate read times will occur.

Choose your country

Americas

Brazil
Brazil
Canada
Canada
United States
United States

Pacific Asia

Australia
Australia
India
India
Japan
Japan
Korea, Republic of
Korea, Republic of
Singapore
Singapore

Europe

Austria
Austria
Belgium
Belgium
Denmark
Denmark
Estonia
Estonia
Finland
Finland
France
France
Germany
Germany
Iceland
Iceland
Italy
Italy
Luxembourg
Luxembourg
Netherlands
Netherlands
Norway
Norway
Poland
Poland
Spain
Spain
Sweden
Sweden
Switzerland
Switzerland
United Kingdom
United Kingdom