Introducing GoTaq® qPCR Master Mix: The Bright Choice for Dye-Based qPCR
The new GoTaq® qPCR Master Mix is a 2X master mix for quantitative PCR (qPCR) and features GoTaq® Hot Start Polymerase, a full-length Taq DNA polymerase bound to a proprietary antibody that prevents polymerase activity at room temperature. The master mix is composed of an optimized buffer formulation complete with dNTPs and MgCl2 and includes a new, proprietary dsDNA-binding dye that can be used at a higher concentration than SYBR® Green I because it is less inhibitory in an amplification reaction. The dye concentration in the master mix is optimized to produce significantly brighter fluorescence during qPCR than master mixes containing SYBR® Green I. Excitation and emission of the dye are similar to those of SYBR® Green I, so it is compatible with commonly available real-time PCR instrumentation.
Karen L. Reece1, Katharine Driftmier Miller Hoffmann1, Cesear Corona2,
Thomas A. Kirkland2, H. Tetsuo Uyeda2, Stephen J. Dwight2, Mark G.
McDougall2 and Douglas R. Storts1
1Promega Corporation, Madison, WI
2Promega Biosciences, LLC., San Luis Obispo, CA
Introduction
Real-time quantitative PCR is a powerful tool to detect and quantify nucleic acids. By incorporating fluorescently labeled probes or nucleic acids or fluorescent double-stranded DNA (dsDNA)-binding dyes into the PCR, product formation can be monitored following each PCR cycle. There are many advantages to using real-time PCR over endpoint PCR. Use of fluorescent dyes to monitor amplification eliminates the need for gel electrophoresis, saving the user time and reducing the potential for contamination. Since product formation can be detected in the exponential phase of PCR where the reaction is most efficient, quantification of starting material is more accurate and sensitive (1). The sensitivity offered by fluorescent dyes also allows detection over a much broader dynamic range than endpoint PCR.
Probe-based chemistries such as TaqMan® technology are advantageous in that the labeled probe, which is specific to the amplicon, produces a fluorescent signal during the enzymatic processing of the probe:amplicon hybrid, indicating the specific synthesis of amplicon. However, these assays are difficult to design and more expensive than dye-based chemistries. dsDNA-binding dyes, such as SYBR® Green I, detect all dsDNA formed during amplification; however, these assays are simple to design and cost-effective. The ability to perform melt curve analysis, which is not possible with the TaqMan® assay, provides a quality control step to verify amplification of the desired amplicon.
GoTaq® qPCR Master Mix
(Cat.# A6001)
introduces a new, proprietary dsDNA-binding dye that can be used at a higher concentration
than SYBR® Green I because it is less inhibitory in an
amplification reaction. The dye concentration in the master mix is
optimized to produce significantly brighter fluorescence during qPCR than
master mixes containing SYBR® Green I. Excitation and emission of the dye are similar to those of SYBR® Green I
"GoTaq® qPCR Master Mix introduces a new, proprietary dsDNA-binding dye that can be used at a higher concentration than SYBR® Green I because it is less inhibitory in an amplification reaction, producing significantly brighter fluorescence. Excitation and emission of the dye are similar to those of SYBR® Green I."
Figure 1. Fluorescence spectra of SYBR® Green I and the proprietary Promega dye in the GoTaq® qPCR Master Mix.
Fluorescence emission spectra were collected for both SYBR® Green I and
the proprietary Promega dye in the absence of dsDNA and in the presence of 50 ng/μl or
Methods
Unless otherwise noted, 50 μl reactions were prepared, and PCR
was performed using an Applied Biosystems 7500 Real Time PCR
instrument in 9600 emulation mode and calibrated according to the
manufacturer’s instructions using the Pure Dye Calibration Kit
(Applied Biosystems). Unless otherwise noted, cycling conditions
were as follows: Activation at 95°C for 2 minutes, 40 cycles of
denaturation at 95°C for 15 seconds and annealing/extension at 60°C for
60 seconds, followed by melt analysis
ramping at 60°C to
Results
Brightness and Sensitivity
The brighter fluorescence signal provided by the GoTaq® qPCR Master Mix enables earlier detection of amplification products, resulting in earlier quantification cycle (Cq) values for many targets compared to SYBR® Green I (Figure 2). This property is demonstrated by amplifying glyceraldehyde-3-phosphate dehydrogenase (GAPDH) from serial dilutions of Human Genomic DNA (Cat.# G3041) using either GoTaq® qPCR Master Mix or Company A’s master mix on the same 96-well plate. The instrument was programmed for an initial 10-minute enzyme activation step at 95°C to accommodate the requirements of Company A’s master mix. The significantly brighter fluorescent signal generated with GoTaq® qPCR Master Mix enables the analysis software to call the Cq values up to three cycles earlier.
Note: Cq is often referred to as cycle threshold or CT. Please refer to Bustin, S. et al. (2009) for recommended qPCR terminology (3).
Figure 2. Performance comparison of GoTaq® qPCR Master Mix and Company A's master mix.
GAPDH was amplified from tenfold serial dilutions
Amplification of a kanamycin phosphotransferase target sequence from plasmid DNA (Figure 3) shows efficient,
linear amplification over
Figure 3. Amplification of kanamycin phosphotransferase from plasmid DNA.
Figure 4. Amplification of Alpha V from cDNA.
Alpha V was amplified from serial dilutions of cDNA
Versatility
GoTaq® qPCR Master Mix is compatible with several real-time PCR instrument platforms using preset filters for SYBR® Green I or FAM™ and ROX™ for the passive reference. Depending on instrument detection settings, however, some instruments require a higher concentration of reference dye in the reaction mix. A tube of 100X CXR Reference Dye is provided with the master mix. Table 1 lists real-time PCR instruments tested and indicates which instruments require additional reference dye.
| Instruments that do not require addition of 100X CXR Reference Dye: |
|---|
| Applied Biosystems 7500 and 7500 Fast Real Time PCR Systems |
| Stratagene Mx3005P® Quantitative PCR Systems |
| Roche LightCycler® 480 |
| BioRad Chromo4™ Real-Time Detector |
| Eppendorf Mastercycler® ep realplex4 and realplex4 S |
| Instruments that require addition of 100X CXR Reference Dye to 1X per reaction: |
| Applied Biosystems 7000 Sequence Detection System |
| Applied Biosystems 7300 Real-Time PCR System |
| Applied Biosystems 7700 Sequence Detection System |
| Applied Biosystems 7900HT Real-Time PCR System |
| Applied Biosystems StepOne® and StepOne®Plus Real-Time PCR Systems |
Bio-Rad real-time instruments, such as the iCycler®, iQ™5 and MyCycler™, perform normalization by collecting Dynamic Well Factors using the background signal detected during the first few PCR cycles. Because the background using SYBR® Green I is not sufficient for the instrument to make these calculations, Bio-Rad protocols require the addition of 10 nM fluorescein per reaction (4). GoTaq® qPCR Master Mix was tested on a Bio-Rad iQ™5 instrument with and without addition of fluorescein. Figure 5 shows that amplification of GAPDH from serial dilutions of human genomic DNA was linear and efficient both with and without addition of fluorescein to the master mix. This suggests that the brightness of the proprietary Promega dye provides sufficient signal for the instrument to collect Dynamic Well Factors without addition of fluorescein. When using a Bio-Rad instrument with GoTaq® qPCR Master Mix, users should run a test reaction using their experimental target to determine if it is necessary to add fluorescein to the master mix.
Figure 5. Performance of GoTaq® qPCR Master Mix using the Bio-Rad iQ™5 with and without fluorescein.
GAPDH was amplified from serial dilutions of human genomic DNA (0.01–100 ng) using the default cycling protocol on the Bio-Rad iQ™5 Real Time instrument. Panel A. Detection of GAPDH without fluorescein. Panel B. Detection with 10 nM fluorescein added (Bio-Rad Cat.# 170-8780). Insets show standard curve data and dissociation profile.
Many companies offer a separate master mix containing an enzyme compatible with fast activation times and faster cycling conditions. GoTaq® qPCR Master Mix is compatible with fast cycling protocols (Figure 6). Reactions (20 μl) were assembled with GAPDH-specific primers and serial dilutions of human genomic DNA. GoTaq® qPCR Master Mix and Company A’s fast master mix were run on separate plates using default fast cycling conditions on an Applied Biosystem 7500 FAST Real Time PCR System (activation at 95°C for 20 seconds, 40 cycles of denaturation at 95°C for 3 seconds and annealing/extension at 60°C for 30 seconds, and melt curve analysis ramping from 60°C to 95°C). GoTaq® qPCR Master Mix performs comparably with Company A’s fast master mix and exhibits a significantly brighter fluorescence signal.
Figure 6. Performance of GoTaq® qPCR Master Mix using a fast cycling protocol.
GAPDH was amplified from serial dilutions of human genomic DNA (0.01–10 ng) using the default fast cycling protocol on the Applied Biosystems 7500 FAST Real Time instrument. GoTaq® qPCR Master Mix is shown on the left, and Company A’s fast master mix is shown on the right. Insets show standard curve data and dissociation profile.
Summary
GoTaq® qPCR Master Mix provides superior amplification and fluorescence qualities in a convenient 2X formulation, which includes a new, proprietary dsDNA-binding dye that results in a significantly brighter signal during qPCR than SYBR® Green I. Since the spectral properties of this dye are similar to SYBR® Green I, the master mix can be easily substituted for other dye-based qPCR master mixes. The convenient 2X master mix requires addition of only experimental template DNA, primers and water. It is compatible with common thermal cycling protocols and fast cycling protocols and can be used on any real-time instrument capable of detecting SYBR® Green I or FAM™ dye.
Reference
- Sundquist, T. (2005) Your key to real-time quantitative PCR. eNotes.
- Knoche, K. and Wheeler, S. (2008) Use the new GoTaq® Hot Start Green and Colorless Master Mixes for easy, convenient PCR. eNotes.
- Bustin, et al. (2009) The MIQE guidelines: Minimum information for publication of quantitative real-time PCR experiments. Clin. Chem. 55, 611–22.
- BioRad iCycler iQ™ Real-Time PCR Detection System Instruction Manual, 4006200 Rev E. This can be viewed online at: www.bio-rad.com/webroot/web/pdf/lsr/literature/4006200E.pdf.
How to Cite This Article
Scientific Style and Format, 7th edition, 2006
Reece, K.L., Hoffmann, K.D.M, Corona, C., Kirkland, T.A., Uyeda, H.T., Dwight, S.J., McDougall, M.G. and Storts, D.R. Introducing GoTaq® qPCR Master Mix: The Bright Choice for Dye-Based qPCR. [Internet] 2009. [cited: year, month, date]; 1, tpub_014. Available from: http://www.promega.com/pubs/tpub_014.htm
American Medical Association, Manual of Style, 10th edition, 2007
Reece, K.L., Hoffmann, K.D.M., Corona, C., Kirkland, T.A., Uyeda, H.T., Dwight, S.J., McDougall, M.G. and Storts, D.R. Introducing GoTaq® qPCR Master Mix: The Bright Choice for Dye-Based qPCR. Promega Corporation Web site. http://www.promega.com/pubs/tpub_014.htm Updated October 2009. Accessed Month Day, Year.
GoTaq is a registered trademark of Promega Corporation. ImProm-II is a trademark of Promega Corporation.
Chromo4, iQ5 and MyCycler are trademarks of Bio-Rad Laboratories. FAM, OneStep is a registered trademark and ROX is a trademark of Applera Corporation. iCycler is a registered trademark of Bio-Rad Laboratories, Inc. LightCycler is a registered trademark of Roche Diagnostics, GmbH. MasterCycler is a registered trademark of Eppendorf-Netheler-Hinz GmbH. Mx3005P is a registered trademark of Stratagene. SYBR is a registered trademark of Molecular Probes, Inc. TaqMan is a registered trademark of Roche Molecular Systems, Inc.
Products may be covered by pending or issued patents or may have certain limitations. Please visit our web site for more information.
All prices and specifications are subject to change without prior notice.
Product claims are subject to change. Please contact Promega Technical Services or access the Promega online catalog for the most up-to-date information on Promega products.