GloMax®-Multi+ Installation and Operational Qualification

Service Performed On-Site by Certified Promega Representative

  • Documentation of instrument function for lab records meets IQ/OQ requirements
  • Easy to use QC worksheet
  • Plate-based instrument testing

Choose an option

Catalog number selected: SA1101

Your price:
Add to Cart
This product is discontinued
GloMax®-Multi+ Installation and Operational Qualification
1 each
Your price: Log in
Change Configuration

Service Options

Installation Qualification

Includes a series of formal instrument checks, delivers written documentation of instrument functionality and demonstrates that everything ordered with the instrument is supplied and installed at the customer’s laboratory. This service product must be delivered by a Promega representative who is certified to perform the Installation Qualification. The service product involves a site visit to perform:

  • installation by qualified Promega personnel
  • inspection of shipping containers, instrument and accessories
  • comparison of items received against items on the purchase order
  • inspection of laboratory conditions
  • review of all hazards and precautions with users
  • confirmation/installation of correct software version
  • instrument test run
  • documentation of Installation Qualification

Operational Qualification

Demonstrates that the instrument functions according to its operational specifications. This service product must be delivered by a Promega representative who is certified to perform the Operational Qualification. The service product involves a site visit to:

  • run operational verification tests
  • document all test results
  • train customer(s) to operate the instrument
  • train customer(s) to use the log book
  • complete Operational Qualification documentation

Specifications

You are viewing: SA1101 Change Configuration

Certificate of Analysis

Search by lot number

Specifications

You are viewing: SA1102 Change Configuration

Certificate of Analysis

Search by lot number

Specifications

You are viewing: SA1103 Change Configuration

Certificate of Analysis

Search by lot number

Let's find the product that meets your needs.

Talk to a Scientist

Alessandro

Alessandro

Italy