Quantus™ Fluorometer Instrument Software and Firmware Download

Download the latest Quantus™ Fluorometer Instrument Software and Firmware Files.

Need help? 

Chat with Technical Service

Software and Firmware Download

Quantus™ Fluorometer Software

The Quantus™ Software is used to transfer data from the Quantus™ Fluorometer to a personal computer.

Software Installation Instructions:

Note: The Quantus™ Software requires the U.S. English locale for data export.

  1. Download the Quantus™ Software above.
  2. Extract the contents of the download file.
  3. Right-click on the “setup” file.
  4. Select “Run as Administrator”.
  5. Follow the prompts to complete the software installation.
  6. If .NET 4.0 framework is not installed on the computer, the software installer will prompt the user to download and install the framework software (available for free from the Microsoft website).

Need help? techserv@promega.com

Quantus™ Fluorometer Firmware

Keep this updated to ensure your Quantus™ Fluorometer works with the latest software releases.

Firmware Installation Instructions:

  1. After downloading the Quantus™ Instrument Firmware and Firmware Loader, extract the contents of the download file.
  2. Press the right and left key pad buttons simultaneously on the Quantus™ Fluorometer, and while pressing these buttons connect the Quantus™ Fluorometer to a USB port on the PC.
  3. From the extracted download file, open the Firmware Loader by double-clicking the “AT91_CDC-USB_FWLoader” file.
  4. Click the “Open FW File” button on the Firmware Loader.
  5. Navigate to the extracted firmware file. This will be called “Quantus – v2.25.bin” or something similar, based on the version.
  6. Click “Open”.
  7. Click the “Update FW” and wait until firmware updates successfully.
  8. Power Off / On the Quantus™ Fluorometer.
  9. Navigate to the Tools screen. The new firmware number should be displayed.