DNA Quantification
Identity Automation™ for the Plexor® HY System setup removes user-to-user variability from qPCR...and cuts hands-on time from an hour or more to just minutes.
- The Plexor® HY System is a real-time PCR assay designed to simultaneously quantify total human and male DNA.
- The Identity Automation™ package provides additional accessories and consumables to allow high-precision setup of the Plexor® HY System.
Performance
- Set up Plexor® HY (80 samples), with dilution of standards, in 30–60 minutes.
- Hands-on time is approximately 10 minutes.
- Amplification mix creation and dilution of standards are fully automated.
- Run precision is equivalent to manual use.
| Total DNA | Male DNA | |||
|---|---|---|---|---|
| Automated | Manual | Automated | Manual | |
| Plexor® HY data were collected for three manual runs and four automated runs on the Biomek® 3000 Workstation. | ||||
| Slope %CV | 1.1% | 3.1% | 2.2% | 1.9% |
| Intercept %CV | 1.9% | 0.4% | 0.5% | 1.9% |
| R2 %CV | 0.2% | 0.2% | 0.2% | 0.5% |
Learn More About the Plexor® HY System
Supported Platforms
- Beckman Coulter Biomek® 3000 Laboratory Automation Workstation
- Beckman Coulter Biomek® NXP Laboratory Automation Workstation
- Tecan Freedom EVO® Workstation